BlueLab Guardian Monitor pH & EC - combined ph, temperature and conductivity tester
-
Code:870
-
Weight:0.600 Kgs
Features:
- "Plant Safe" green LED display
- Visual alarm for high and low settings
- Large, easy-to-read displays
- Greater tolerance to RF/electronic interference is provided
- No conductivity or temperature calibration required
- Waterproof design
- Selectable conductivity and temperature units
- International power supply
- Easy pH calibration at the push of a button
- Adjustable display brightness
- Interchangeable dual compound pH probe
- Settings are saved in case of power loss (non-volatile memory)
How to set up/mount and calibrate, see below in description
Quick Buy
BlueLab Guardian Monitor
The Bluelab Guardian Monitor is a continuous indicator of the current EC, pH and temperature levels of your nutrient solution, allowing you to optimize these parameters as your plants progress through each growing phase.
How to install:
1. Mount the unit on the wall or hang it on a support to ensure that the probes can reach the tank.
2. Connect the appropriate plug adapter to the power supply.
3. Connect the power supply and the Bluelab pH probe to the base of the Guardian Monitor.
4. Turn on the mains power.
How to set up:
Change the conductivity and temperature units
1. Press and hold the conductivity units button. After 3 seconds it will move to the next unit indicated by the glowing LED.
Release the button. Repeat the step until you select the desired unit.
2. Press and hold the temperature units button. After 3 seconds it will move to the next unit indicated by the glowing LED.
Release the button. Repeat the step until you select the desired unit.
Select display brightness
1. Press and hold the "Display Brightness" button while pressing the up or down buttons to change the brightness.
How to calibrate:
Calibration of the pH probe must be performed before use . The EC probe does not require calibration, but both probes must be in the pH solution to compensate for temperature.
1. Remove the storage cap from the pH probe with a slight twist.
2. Rinse the tip of both probes with fresh water, shake off excess water.
3. Place the pH and EC probe in Bluelab pH 7.0 calibration solution.
4. Press and hold the "pH cal" button until "PH" and "CAL" appear. Release the button. The pH 7 LED and the first indicator will blink, increasing to four while the meter is calibrating. The pH 7 LED will light up when complete.
5. Rinse both probes in fresh water, shake off excess water.
6. Place the tip of both probes in Bluelab pH 4.0 Calibration Solution or pH 10.0 Calibration Solution.
7. Press and hold the 'pH cal' button until 'PH CAL' appears. Release button. The pH 4/10 LED and the first indicator will flash, increasing to four as the device calibrates. When finished, the pH 4/10 LED will light up. Calibration is complete.
8. If you do not place the pH probe in the tank/reservoir immediately, replace the storage cap. The tip of the pH probe should not be allowed to dry out . Hydrate with Bluelab KCl Storage Solution if needed.
After 30 days the LEDs will flash indicating that the pH probe needs calibration.
Suitable calibration solutions can be found here here , here , here and here
You can view a spare ph probe here
CAUTION: Do not pour concentrated nutrient liquids or pH correctors directly onto probes when submerged!
The tester is guaranteed against defects for a period of 2 years from the date of purchase. During this period, the device can be returned to the seller with proof of purchase (receipt). If a defect is found, it will be repaired or replaced.
This warranty does NOT cover misuse of the device, accidental damage (abrasion, getting wet), repairs by unauthorized persons. 6 month warranty on the Ph probe.
Found a better price?
Delivery conditions
CONDITIONS FOR DELIVERY
- Door-to-door orders on the territory of Bulgaria are sent with Econt Express on the next day of the order.
- Delivery to other countries is estimated within 3-6 business days.
- If the item you ordered is not available, we will contact you to inform you about the delivery time and to confirm your order.
- The delivery price is indicative and is recalculated according to the rates of the courier company.
- All goods are packed in a way that prevents any damage. If you notice the packaging is damaged, once your package has arrived, inform the transporter!
- Goods are always packed in discreet packaging.
NOTE: The delivery companies may experience delays on Black Friday and Official Holidays due to an increase in shipping volume. Thank you for your patience and understanding.
Order items at a regular price with a total value of over BGN 100 and pick up with FREE DELIVERY to an Econt office.